Change the position of the transducer
Change the position of the transducer
Change the position of the transducer
Change the position of the transducer
•
Unlock the floating transducer
2
by turning the locking knob
1
counter-clockwise.
•
Slide the transducer
2
to the advised mounting distance
3
(menu X7.2.3).
•
Lock the transducer by turning the locking knob
1
clockwise.
Greasing the transducer surfaces
Greasing the transducer surfaces
Greasing the transducer surfaces
Greasing the transducer surfaces
•
1
press in both locking strips on the left and right end of the sensor rail.
•
2
lift the cover in vertical movement, then tilt over the cover 90
°
3
.
•
4
put grease on the contact surfaces of the transducers.
•
5
put back cover by tilting back in a 90
°
angle.
•
6
press the cover back vertically on the locking strips until you hear them click.