Mounting the DOWN rail in Z-mode
Mounting the DOWN rail in Z-mode
Mounting the DOWN rail in Z-mode
Mounting the DOWN rail in Z-mode
Measure the circumference of the pipe with a measuring band.
For Z-mode, you must install the DOWN rail at the opposite location at the pipe. There are two
possible ways to find the exact location:
Find the location with a fixed reference point
Calculate the half of the circumference. Mark this 180
°
alignment line on the pipe.
•
Mount the DOWN rail in such a way that the transducer is at the marked location.
Figure 4-5: Find the opposite location with a reference point
1
Measure the distance between the transducer of the UP rail and the reference point.
2
Add the Advised Distance and mark the location on the alignment line.