ELECTRICAL CONNECTIONS
3
37
OPTISONIC 6300
www.krohne.com
06/2019 - 4005857501 - QS OPTISONIC 6300 v2 R01 en
3.6 Signal cable to converter
The flow sensor is connected to the signal converter via one signal cable, with (labelled) inner
coax cables for the connection of the acoustic paths.
INFORMATION!
Connect the cable to the connector with similar numeral marking.
Field version
Figure 3-10: Connect signal cable
Construction of console (F-version)
Figure 3-11: Inserting cable and secure with clamp on shielding bush
1
Cables
2
Cable glands
3
Grounding clamps
4
Cable with metal shielding bush
CAUTION!
Re-connecting of the coax connectors is limited. Make sure that the male connector on the coax
cable, is always put straight on the female connector in the connection terminal of the unit.
Excessive dis- / re-connection and/or positioning the connectors skewed to each other will
damage the inside clips of the connectors. This results in an improper contact and measurement
errors.