172
·
Adjustment
Operation procedure
Key
Display
⑴
After levelling the machine, power on, press
CNFG under date/ time screen, as right diagram
shows.
CNFG
2007-09-10 10
:
00
:
48
Type
:
KTS
-
440R
No.
:
S12926
Ver.
:
07.09.10
Job
:
JOB01
MEAS MEM CNFG
⑵
Press ▼ key to choose“2. Instr. Const.”and press
ENT key(or press numeric key 2),enter the
instrument constant setting screen.
“
2
. Instr.
Const.”
+
ENT
1. Obs. condition
2. Instr. Const.
3. Date& time
4. Comms setup
5. Unit
6. Key function
⑶
press ▼ key to choose “2.V0/ Adjustment”, then
press ENT key(or press numeric key2), enter
collimation error adjustment function.
“2. V0/
Adjustment”
+
ENT
1. V0/Axis const.
2. V0/ Adjustment
3. Collimation
4. horizontal axis
5. Instr. Const.
6. Contrast ADJ.
⑷
At positive position (Face left) sight at target,
press OK.
Face left,
sight at target
+
OK
V0 adjustment
<Step-1> Front
ZA 0°21′49″
HAR 185°47′42″
OK
⑸
Rotate telescope. At the reverse position (face
right) sight at the same target precisely, press OK.
Face right
and sight at
target
+
OK
V0 Adjustment
<Step- 2> Reverse
ZA 179°38′30″
HAR 5°50′15″
OK
⑹
After adjustment, it shows “set” on the screen,
the display returns to Instrument constant menu
screen.
1. V0/Axis const.
2. V0/ Adjustment
3. Collimation
4. horizontal axis
5. Instr. Const.
6. Contrast ADJ.
4
Repeat the inspection steps to measure the Index Difference (
i
angle). If the Index
Difference can not meet the requirement, you should check whether the three steps of the Adjustment are right, the
sight is right and etc. Then set again according to the requirement.
5
If Index Difference can not still meet the requirement after the repeated operation, the instrument should be
returned to factory for inspection and repair.