32
3.5.2 Connecting an Analog Sensor/Second Memosens Channel
CAUTION! Electrostatic discharge (ESD).
The modules’ signal inputs are sensitive to electrostatic
discharge. Take measures to protect against ESD before inserting the module and wiring the inputs.
NOTICE!
Strip the insulation from the wires using a suitable tool to prevent damage.
Stripping length
Measuring Modules for Connection of Analog Sensors: pH, ORP, Oxygen, Conductivity
1
2
3
4
01.
Switch off the power supply to the device.
02.
Open the device (loosen the 4 screws on the front).
03.
Loosen screw
(1)
on the module cover
(2)
(“ESD shield”) and open the cover.
04.
Insert the module into the module slot
(3)
.
05.
Attach the module plate sticker
(4)
.
06.
Connect the sensor and separate temperature probe, if necessary.
Channel II Wiring Examples, p. 195
CAUTION! Risk of losing the specified ingress protection.
Fasten the cable glands and screw
together the housing correctly. Observe the permissible cable diameters and tightening torques.
Only use original accessories and spare parts.
07.
Check whether all connections are correctly wired.
08.
Close the module cover
(2)
, tighten screw
(1)
.
09.
Close the device and tighten the screws on the front.
10.
Switch on the power supply.
Now select a measuring function and configure the sensor
01.
From within measuring mode, press the
left softkey: Menu
.
✓
The
Menu Selection
opens.
02.
Select
Parameter Setting
Sensor Selection [I] [II]
.
Sensor Selection
Sensor Selection
Back
Sensor Selection
Lock
Power Out
Off
I
II
(Admin.)
I
II
Note:
Function check (HOLD) is active.
03.
Press
enter
to open the
Sensor Selection [II]
.