Stratos Multi E461N
139
05.
Save with the
right softkey: Save
.
✓
An information window is shown.
06.
Right softkey: Close
07.
As required, exit calibration by pressing the
left softkey: Back
.
Note:
The icon
indicates that product calibration has not yet been completed.
Step 2: Lab value has been measured.
08.
Open the Product Calibration menu again.
Calibration
Calibration Mode
Back
Product
(Step 2: Lab Value)
Proceed
II
09.
Right softkey: Proceed
10.
Enter the lab value and press
enter
to confirm.
11.
Confirm with the
right softkey: Next
or repeat calibration with the
left softkey: Cancel
.
Result
✔
The calibration record is displayed. By pressing the
right softkey: Adjust
, the calibration values
obtained during calibration for calculation of the process variables are applied to the device.
When using a Memosens sensor, the calibration values are saved in the sensor.
Exception: Sample value can be determined and entered on site:
12.
Take sample.
✓
The measured value and temperature at the time of sampling are displayed.
13.
Left softkey: Entry
14.
Enter the lab value and press
enter
to confirm.
15.
Confirm with the
right softkey: Next
or repeat calibration with the
left softkey: Cancel
.
Result
✔
The calibration record is displayed. By pressing the
right softkey: Adjust
, the calibration values
obtained during calibration for calculation of the process variables are applied to the device.
When using a Memosens sensor, the calibration values are saved in the sensor.