Stratos Multi E461N
153
Sensor Wear Monitor
Note:
Function active for digital sensors.
The sensor wear monitor shows the sensor operating time and the maximum temperature during
the operating time, as well as wear and the forecast remaining time. When using oxygen sensors,
the number of membrane replacements and calibrations is also displayed:
Diagnostics
[I] [II] [Sensor]
Sensor Wear Monitor
Sensor Wear Monitor
Back
Operating Time
68 d
Back to Meas.
I
Wear
Remaining Lifetime
Max. Temperature
9.5 %
661 d
32 °C
Load Matrix
Note:
Function active for ISM pH/ORP sensors and amperometric ISM oxygen sensors.
Back
Sensor 1 Load Matrix
<2
2 ... 4
4 ... 8
8 ...
11
>11
<10
°C
>80
pH
Bar color
Green:
The area that puts the least load on the sensor.
Yellow:
The area that puts more load on the sensor.
Red:
The area that puts the most load on the sensor.
The bar height indicates the duration of the load.
See also
Digital ISM Sensors (FW-E053), p. 241
Statistics
Note:
Function active for ISM pH/ORP sensors and amperometric ISM oxygen sensors.
The statistics data provides information on the sensor product life cycle: Data from the first adjustment
and the three most recent calibrations/adjustments is displayed. This data can be used to assess the
performance of the sensor over its service life.
Use the
right softkey
to switch between a graphical display and a list.
See also