Stratos Multi E461N
20
3.1.2 Housing Mounting Options
Note:
All dimensions are given in millimeters [inches].
Knockouts in the rear unit enable different mounting options:
• Wall mounting
• Pipe mounting
• Panel mounting
• Protective hood
Protective Hood for Wall and Pipe Mounting ZU0737, p. 22
Cable glands for connecting sensors:
• 3 knockouts for M20x1.5 cable glands
Blanking Plugs, Reduction Sealing Inserts, Multiple Sealing Inserts, p. 24
• 2 knockouts for M20x1.5 or NPT 1/2
″
cable glands or rigid metallic conduit
Note:
Install the hinge pin to prevent tensile strain on the measuring cables when replacing the front
unit. Imprecise measuring values may result if this is not done.
Mounting Clearance
110
[4.33]
110
[4.33]
There is a 100 mm hinge pin in the bag containing small accessory parts included in the package
contents
Package Contents and Product Identification, p. 15
. The hinge pin connects the front and
rear units. Depending on space requirements, the hinge pin can be inserted on the left or right. In
order to replace the front unit, a minimum clearance of 110 mm [4.33 inches] must be maintained on
the relevant side.