Stratos Multi E461N
39
5.3 Menu Structure Overview
Measuring
2958
1989 Administrator Level
1246 Operator Level
Passcode:
(factory
setting)
Calibration
Maintenance
Parameter Setting
Diagnostics
Menu
1147
Message List, Logbook,
Network Information,
Monitor, Device/Sensor
Information
System Control (Time, FW Update, Options,
Passcode Entry), Sensor Selection, General
(Language, Measurement Display), Inputs/
Outputs, Measuring Channels I and II, PROFINET
Maintenance of
sensors and current
outputs
Calibration mode,
calibration points,
buffer set
1
2
3
4
5
6
Submenus:
1
Pressing the
left softkey: Menu
accesses the Menu
Selection.
4
Press
enter
to confirm, enter passcode.
2
Pressing the
right softkey: Back to Meas.
returns to
measurement.
5
Further submenus and menu items are shown.
3
Select a menu using the
arrow keys
.
6
Selected functions in the Diagnostics menu can
also be accessed in measuring mode by pressing
the right
softkey
(Favorites menu).
5.4 Access Control
Access to the device functions is regulated and limited by individually adjustable passcodes. This pre-
vents unauthorized modification of device settings or manipulation of the measurement results.
Passcodes are configured in
Parameter Setting
System Control
5.5 Operating States
Function Check Mode (HOLD Function)
After activating parameter setting, calibration, or maintenance, the Stratos Multi enters function check
mode (HOLD). PROFINET communication and the relay contacts/current outputs behave in accor-
dance with the parameter settings. The state transmitted via PROFINET is in part dependent on the
operating mode.