Stratos Multi E461N
67
Adjusting Sensor Monitoring
01.
Sensor Data
Sensor Monitoring Details
.
02.
Open a sensor parameter, e.g.,
Slope
.
03.
Set
Monitoring
of the slope to automatic or individual.
04.
If you select “Individual”: The nominal slope and the min/max limits can be entered.
05.
In the
Message
menu item, select whether and how an exceeded limit is to be displayed:
Off
No message, but the parameter is still shown in the Diagnostics menu and on the sensor diagram.
Failure
A Failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
If “Display Color NE107” is selected, the measurement display is shown with red backlighting.
Maintenance
A Maintenance Required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed. If “Display Color NE107” is selected, the measurement display is shown with blue
backlighting.
06.
Set the sensor monitoring details for other sensor data, e.g., zero point, Sensocheck, response
time, sensor wear, or sensor operating time.
07.
Using the
left softkey: Back
, confirm the sensor monitoring settings and adjust further parameters.
or
Using the
right softkey: Back to Meas.
, confirm the sensor monitoring settings and end the
function check (HOLD).
CIP/SIP Counters
CIP/SIP counters are available for the following pH sensor types:
Memosens pH
Memosens pH/ORP
ISM pH/ORP
1)
CIP counter
+
+
SIP counter
+
+
+
CIP/SIP cycles are used to clean or sterilize process-wetted parts in the process. Depending on the
application, either one chemical (alkaline solution, water) or several chemicals (alkaline solution,
water, acidic solution, water) are used.
• CIP temperature > 55 °C/131 °F
• SIP temperature > 115 °C/239 °F
The cleaning (cleaning in place) and sterilization (sterilization in place) cycles are counted to measure
the load on the sensor, e.g., in biotechnology applications.
Note:
If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a Maintenance Required message is triggered and the
NAMUR icon
is displayed; the measurement display is shown with blue backlighting (display color:
NE107).
Note:
A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note:
With Memosens sensors, an entry is also made in the sensor.
1)
With TAN option FW-E053