Stratos Multi E461N
83
6.11 Conductivity (Inductive) Process Variable
Note:
Function check (HOLD) is active.
Note:
After changing the process variable or measuring mode, Stratos Multi retains its settings but
needs to be reconfigured.
Selecting a Digital Toroidal Conductivity Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [I]
Selection of a Memosens toroidal conductivity sensor connected to the RS-485 interface
(terminals 1 … 5):
Process variable:
Auto or Conductivity (Induct.)
Mode:
Memosens
Functionality:
CondI
Selection of an SE670/SE680K toroidal digital conductivity sensor connected to the RS-485 interface
(terminals 1 … 5):
Process variable:
Conductivity (Induct.)
Mode:
Other digital
Functionality:
SE670/SE680K
Selecting a Second Digital Toroidal Conductivity Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [II]
Selection of a second Memosens conductivity sensor connected to the MK-MS095N measuring
module:
Module:
MK-MS
Process variable:
Conductivity (Induct.)
Mode:
Memosens
Functionality:
CondI
Selection of a second SE670/SE680K digital toroidal conductivity sensor connected to the MK-MS095N
measuring module:
Module:
MK-MS
Process variable:
Conductivity (Induct.)
Mode:
Other Digital
Adjustable Parameters for Digital or Memosens Sensors for Inductive Conductivity
Parameter Setting
[I] [II] Digital/Memosens CondI
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
Sensoface
Enable/disable display of the Sensoface icon.
If “Other Digital” is selected:
Sensocheck
Monitoring of primary and secondary coils.
Enable or disable Sensocheck to generate Failure or
Maintenance Required messages.
If Memosens is selected:
Sensor Monitoring Details
Option to enter individual limits for monitoring the cell factor.
Sensocheck: Monitoring of primary and secondary coils.
Enable or disable Sensocheck to generate Failure or
Maintenance Required messages.
Option to enter individual limits before a message for SIP
counter, CIP counter, and sensor operating time is triggered.
Cal Presettings
Presetting of calibration mode with corresponding parameters.
Calibration Presettings, p. 87
TC Process Medium
Temperature Compensation of Process Medium, p. 88
Concentration