Stratos Multi E461N
159
Error No. Type Message Text
Notes/Solution
F200
Configuration Data
Loss
Data error in the configuration: Reset to factory settings and fully
reconfigure.
F202
System Failure
Internal system error: Switch device off (approx. 10 s). If the message
persists, send in the device.
F203
Inconsistent Parame-
ter Setting
The configuration of the measuring channel operating mode is
inconsistent: Check and correct the configuration.
F210
Device Diagnostics:
Self Test
Error on the self test (RAM): Switch device off (approx. 10 s). If the message
persists, send in the device.
F212
Time/Date
The time and date still need to be set.
Parameter Setting
System Control
Time/Date
F229
Info
Wrong Passcode
An incorrect passcode was entered.
F234
Info
Key Lock is Active
Disable key lock via PCS: Slot 1, subslot 1, index 1
Current Output/Relay Contacts
Error No. Type Message Text
Notes/Solution
B070
Current I1 Span
Current output 1: The selected span is too small/too big:
Parameter Setting
Inputs/Outputs
Current Outputs
Current Output I1
Check start/end.
B071
Current I1 < 0/4 mA
Current output 1: The current is below the permissible limit.
B072
Current I1 > 20 mA
Current output 1: The current is above the permissible limit.
B073
Current I1 Load Error
Current output 1: The current loop has been interrupted (open circuit) or
the load is too high: Deactivate or short-circuit unused current outputs.
B074
Current I1 Parameter
Current output 1: Check the configuration.
B075
Current I2 Span
Current output 2: The selected span is too small/too big:
Parameter Setting
Inputs/Outputs
Current Outputs
Current Output I2
Check start/end.
B076
Current I2 < 0/4 mA
Current output 2: The current is below the permissible limit.
B077
Current I2 > 20 mA
Current output 2: The current is above the permissible limit.
B078
Current I2 Load Error
Current output 2: The current loop has been interrupted (open circuit) or
the load is too high: Deactivate or short-circuit unused current outputs.
B079
Current I2 Parameter
Current output 2: Check the configuration.
pH, ORP
Error No. Type Message Text
Notes/Solution
P001
Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
P008
Adjustment Data
Error in the adjustment data:
Switch device off (approx. 10 s).
If the message persists, send in the device.
P009
Firmware Error
Error in the firmware:
Switch device off (approx. 10 s).
Reload the firmware.
Firmware Update (FW-E106), p. 245
If the message persists, send in the device.
P010
pH Range
Measuring range exceeded. Possible causes: Sensor not or incorrectly con-
nected, sensor faulty, cable faulty, incorrect temperature detector selected,
temperature detector faulty.
With MK-PH015 module without connection to solution ground: No jumper
between terminals B and C.
pH Analog Wiring Examples, p. 195
P011
pH LO_LO
Value below configured monitoring limit