Stratos Multi E461N
187
pH calibration and adjustment
Calibration with automatic buffer recognition (Calimatic)
Manual calibration with entry of individual buffer values
Product calibration
Data entry of premeasured sensors
ISFET zero point (with ISFET sensors)
Temperature probe adjustment
Calculation of nominal zero point
Max. calibration range
Asymmetry potential (zero point): ±60 mV
Slope: 80 ... 103 % (47.5 ... 61 mV/pH)
Zero adjustment
±750 mV for Memosens ISFET
Buffer sets
Knick CaliMat
2.00/4.00/7.00/9.00/12.00
Mettler-Toledo
2.00/4.01/7.00/9.21
Merck/Riedel
2.00/4.00/7.00/9.00/12.00
DIN 19267
1.09/4.65/6.79/9.23/12.75
NIST Standard
1.679/4.005/6.865/9.180
NIST technical
1.68/4.00/7.00/10.01/12.46
Hamilton
2.00/4.01/7.00/10.01/12.00
Kraft
2.00/4.00/7.00/9.00/11.00
Hamilton A
2.00/4.01/7.00/9.00/11.00
Hamilton B
2.00/4.01/6.00/9.00/11.00
HACH
4.01/7.00/10.01
Ciba (94)
2.06/4.00/7.00/10.00
WTW techn. buffers
2.00/4.01/7.00/10.00
Reagecon
2.00/4.00/7.00/9.00/12.00
Specifiable buffer set
TAN Option FW-E002
ORP calibration and adjustment
ORP data entry
ORP adjustment
ORP check
Temperature probe adjustment
Max. calibration range
-700 ... 700 ΔmV
Adaptive calibration timer
Interval
0000 ... 9999 h