6.14 PROFINET
6.14.1 System Integration
Initial commissioning
PROFINET devices are identified in the network using the following parameters:
IP address, MAC address, and device name (name of station).
• IP address on delivery: 0.0.0.0
• For the MAC address, see the terminal plate.
• Device name (name of station) not set on delivery (empty).
IP configuration and device name (name of station) must be set using a suitable engineering tool.
Configuration with the Siemens TIA Portal is specified on the following pages.
Application
Tool
Integration in PROFINET system
Engineering tool
Device name and IP address
Engineering tool
Device parameter setting
Local device
Back up/save device configuration
Data Card ZU1080-S-N-D
Installing the GSDML File
A PROFINET device master file (GSDML file) is required for system integration.
The latest version of the GSDML file is available in the downloads section of the Knick website.
The procedure for installing the GSDML file varies depending on the engineering tool used. The
Simatic TIA Portal V15 is used in the example.
01.
Open the
Extras
Manage General Station Description Files (GSD)
menu.
02.
Browse
button: Select the directory containing the GSDML file to be installed.
03.
Select the GSDML file from the list.
04.
Install
button: The GSDML file is installed and added to the hardware catalog.