6.15 Flow
Stratos Multi can calculate flow for limit messages or to monitor an ion exchanger. A pulse generator is
connected to control input OK1 for this purpose.
Parameter Setting
Control input OK1 must first be assigned the “Flow” function.
01.
System Control
Function Control
02.
Input OK1
: Select “Flow”.
03.
2x
left softkey: Back
04.
Inputs/Outputs
Control Inputs
Flow
05.
Enter the number of pulses per liter.
06.
If required, enable monitoring of the minimum and maximum flow.
The flow measurement can process up to 100 pulses per second at the signal input of control input
OK1.
Flow Monitoring with a Connected External Flow Transmitter
Factory setting to generate a failure message
Minimum flow
5 liters/h
Maximum flow
25 liters/h
The flow messages can activate a relay contact and/or trigger a 22-mA message via a current output
(user-defined).