171
Oxy/Oxy Calculation Block
Error No. Type Message Text
Notes/Solution
H001
Configuration Data Loss
Data error in the configuration:
Fully reconfigure the device.
H010
Saturation %Air Diff Range Saturation value difference. Max/min device limits exceeded:
- Check both saturation values.
- Check sensor/cable connections.
H011
Saturation %Air Diff LO_LO Value below configured monitoring limit
H012
Saturation %Air Diff LO
Value below configured monitoring limit
H013
Saturation %Air Diff HI
Value above configured monitoring limit
H014
Saturation %Air Diff HI_HI
Value above configured monitoring limit
H015
Temperature Diff Range
Temperature value difference. Max/min device limits exceeded:
- Check both temperature values.
- Check sensor/cable connections.
H016
Temperature Diff LO_LO
Value below configured monitoring limit
H017
Temperature Diff LO
Value below configured monitoring limit
H018
Temperature Diff HI
Value above configured monitoring limit
H019
Temperature Diff HI_HI
Value above configured monitoring limit
H020
Conc. (Liquid) Diff Range
Concentration value difference. Max/min device limits exceeded:
- Check both concentration values.
- Check sensor/cable connections.
H021
Conc. (Liquid) Diff LO_LO
Value below configured monitoring limit
H022
Conc. (Liquid) Diff LO
Value below configured monitoring limit
H023
Conc. (Liquid) Diff HI
Value above configured monitoring limit
H024
Conc. (Liquid) Diff HI_HI
Value above configured monitoring limit
H045
Saturation %O2 Diff Range Saturation value difference. Max/min device limits exceeded:
- Check both saturation values.
- Check sensor/cable connections.
H046
Saturation %O2 Diff LO_LO Value below configured monitoring limit
H047
Saturation %O2 Diff LO
Value below configured monitoring limit
H048
Saturation %O2 Diff HI
Value above configured monitoring limit
H049
Saturation %O2 Diff HI_HI
Value above configured monitoring limit
H090
Conc. (Gas) Diff Range
Concentration value difference. Max/min device limits exceeded:
- Check both concentration values.
- Check sensor/cable connections.
H091
Conc. (Gas) Diff LO_LO
Value below configured monitoring limit
H092
Conc. (Gas) Diff LO
Value below configured monitoring limit
H093
Conc. (Gas) Diff HI
Value above configured monitoring limit
H094
Conc. (Gas) Diff HI_HI
Value above configured monitoring limit
H200
Param. Calculation Block
Check the configuration:
Parameter Setting
System Control
Calculation Blocks