Stratos Multi E461N
77
Selecting an Analog Conductivity Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [II]
Selection of a conductivity sensor connected to the MK-COND025N measuring module:
Module:
MK-COND
Mode:
Analog
Adjustable Parameters for Analog Conductivity Sensors
Parameter Setting
[II] Analog Cond
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
Sensor Type
Select the used sensor type: 2‑electrode sensor, 4‑electrode
sensor, SE600, SE602, SE603, SE604, SE610, SE620, SE630.
Nom. Cell Constant
Enter when 2‑el. or 4‑el. sensor is selected.
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensocheck
Enable or disable Sensocheck to generate failure or maintenance
required messages.
Temperature Detection
Set measuring and calibration temperature.
When 2‑el. or 4‑el. sensor is selected: Select temperature detector.
Cal Presettings
Presetting of calibration mode with corresponding parameters.
Calibration Presettings, p. 80
TC Process Medium
Temperature Compensation of Process Medium, p. 80
Concentration
Concentration (TAN Option FW-E009), p. 81
TDS
Enable/disable the TDS function
USP
Enable/disable USP function for monitoring ultrapure water and set the USP limit.
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
6.10.1 Sensor Data
Memosens sensors provide relevant sensor data automatically.
The sensor type must be selected if using analog sensors:
Parameter Setting
[II] Analog …
Sensor Data
Sensor Data
Sensor Type
2-El. Sensor
Back
1.000 /cm
Nom. Cell Constant
II
Back to Meas.
Temperature Detection
On
Sensoface
Off
Sensocheck
01.
Select the
Sensor Type
.
02.
Enter the sensor’s nominal cell constant.
03.
In
Temperature Detection
, select the used temperature probe and whether the temperature is to
be measured automatically or manually during measurement and/or calibration.