Stratos Multi E461N
182
13 Specifications
13.1 Power
Power supply,
terminals 17, 18
80 V (- 15 %) ... 230 (+ 10 %) V AC; approx. 15 VA; 45 … 65 Hz
24 V (- 15 %) … 60 (+ 10 %) V DC; 10 W
Overvoltage category II, protection class II, pollution degree 2
Test voltage
Type test 3 kV AC 1 min after moisture pre-treatment
Routine test 1.4 kV for 2 s
13.2 Inputs and Outputs (SELV, PELV)
S
ensor input 1
for Memosens/optical sensors (SE 740), galvanically isolated
Data in/out
Asynchronous interface, RS-485, 9600/19200 Bd
Power supply
3.08 V (3.02 … 3.22 V)/10 mA, R
i
< 1 Ω, short-circuit-proof
S
ensor input 2
For measuring module or analog/ISM
1)
measuring module,
galvanically isolated
Data in/out
Asynchronous interface RS-485, 9600 Bd
Power supply
3.08 V (3.02
…
3.22 V)/6 mA, R
i
< 1 Ω, short-circuit-proof
Input OK1
Galvanically isolated (optocoupler)
Switching between parameter sets A/B, flow measurement, function check
Parameter set selection
Relay input 0 … 2 V (AC/DC) parameter set A
Relay input 10 … 30 V (AC/DC) parameter set B
Control current 5 mA
Flow
Pulse input for flow measurement 0 … 100 pulses per second
Display: 00.0 … 99.9 l/h
Message via 22 mA or relay contact
Power out
Power output, short-circuit-proof, 0.5 W, for operating the SE 740 sensor
Off; 3.1 V (2.99 … 3.25 V); 14 V (12.0 … 16.0 V); 24 V (23.5 … 24.9 V)
Output 1, 2
Out 1, Out 2
0/4 ... 20 mA, floating, load resistance up to 500 Ω, galvanically connected
When using the current outputs, neither PROFINET nor the relay contacts can be used.
Failure message
3.6 mA or 22 mA, adjustable
Active
Max. 11 V
Passive
Supply voltage 3 … 24 V
Process variable
Selection from all available process variables
Start/end of scale
Configurable within selected range
Characteristic
Linear, bi-/trilinear, or logarithmic
Output filter
PT1 filter, filter time constant 0 ... 120 s
Measurement error
2)
< 0.25 % of current value + 0.025 mA
Contact REL1, REL2
Relay contact, floating
Contact rating
with ohmic load
AC < 30 V
rms
/ < 15 VA
DC < 30 V / < 15 W
Max. switching current
3 A, max. 25 ms
Max. continuous current
500 mA
User-definable: Failure, maintenance required, function check, min/max limit, rinse contact, parameter set B
signaling, USP output, Sensoface
1)
ISM with TAN option FW-E053
2)
At rated operating conditions