Messages

No.

OXY messages

Message type

D045

Saturation %O2 Range

FAIL

D046

Saturation %O2 Alarm LO_LO

FAIL

D047

Saturation %O2 Alarm LO

WARN

D048

Saturation %O2 Alarm HI

WARN

D049

Saturation %O2 Alarm HI_HI

FAIL

D060

Sad Sensoface: Slope

WARN

D061

Sad Sensoface: Zero

WARN

D062

Sad Sensoface: Sensocheck

User-defined

D063

Sad Sensoface: Response time

WARN

D064

Sad Sensoface: Cal timer

WARN

D070

Sad Sensoface: Sensor wear

User-defined

D071

Sad Sensoface: Membrane wear

User-defined

D080

Range (sensor current)

WARN

D081

O2 measurement OFF (temp)

Text

D090

Vol% range (measurement in gases)

WARN

D091

Vol% Alarm LO_LO (measurement in gases)

FAIL

D092

Vol% Alarm LO (measurement in gases)

WARN

D093

Vol% Alarm HI (measurement in gases)

WARN

D094

Vol% Alarm HI_HI (measurement in gases)

FAIL

D095

ppm range (measurement in gases)

FAIL

D096

ppm Alarm LO_LO (measurement in gases)

FAIL

D097

ppm Alarm LO (measurement in gases)

WARN

D098

ppm Alarm HI (measurement in gases)

WARN

D099

ppm Alarm HI_HI (measurement in gases)

FAIL

D100

No sensor

FAIL

D101

Communication interrupted

FAIL

D102

Sensor connection

FAIL

Summary of Contents for Protos LDO 3400-170

Page 1: ...eitung Messmodul Protos CONDI 3400 X 051 zur Leitfähigkeitsmessung mit induktiven Sensoren Deutsch Protos LDO 4400 170 LDO 3400 170 Measuring Module For Optical Oxygen Measurement in Liquids and Gases User Manual Protos II 4400 X Protos 3400 X Process Analysis System Latest Product Information www knick de ...

Page 2: ...ntact our Service Team before returning a defective device Ship the cleaned device to the address you have been given If the device has been in contact with process medium it must be decontaminated disinfected before shipment In this case place a Declaration of Contamination in the consignment to prevent any risk to the health and safety of our service personnel The declaration is available at htt...

Page 3: ...omatic Calibration in Air 16 Product Calibration Calibration with Sampling 18 Zero Correction 20 Parameter Setting 21 Parameter Setting Operating Levels 22 Administrator Level 22 Operator Level 22 Viewing Level 22 Parameter Setting Locking a Function 23 Activating Parameter Setting 24 Setting the Sensor Data Parameters 25 Sensoface 29 Logbook Factory Setting 30 Setting the Message Parameters 31 Cu...

Page 4: ... Description 38 FRONT Module 38 BASE Module 38 Sensor Monitor 39 Cal Record 40 Sensor Diagram 40 CIP SIP Cycles Autoclaving Counter 41 Setting Diagnostics Messages as Favorite 42 Message List 44 Messages 45 Specifications 52 Appendix 55 Minimum Spans for Current Outputs 55 Overview 56 Overview of Parameter Setting 56 Index 60 ...

Page 5: ...dules against electrostatic discharge Before commissioning it must be proved that the device may be connected with other equipment Safety Information The module is used to measure oxygen in liquids and gases using an SE 740 series optical sensor It measures the partial pressure of oxygen air pressure and temperature simultaneously It is also able to calculate and display the oxy gen saturation and...

Page 6: ...ring mode Press menu key open Diagnostics menu Device Description Sensor Device description Return 22 7 C 83 3 Air Digital sensors Module LDO 3400 170 Hardware 1 Software 2 0 Serial number 471101149 BASE FRONT Sensor Query module firmware Here LDO 3400 170 module hardware and firmware version serial number here installed in slot 3 Module Note The display may vary depending on the device version Mo...

Page 7: ...ol LDO Input 0 4 to 20 mA Input 87708 0000000 jjww http www knick de knick knick de In et tern Made in Germany Tamb 20 to 55 C LDO LDO 3400 170 2 3 4 5 6 7 7 8 8 9 9 10 10 11 11 12 12 13 13 14 14 15 15 16 16 17 17 18 18 19 19 Shield Power Out N C N C A B Control Control GND N C N C Control Control Input 0 4 to 20 mA Input 0 4 to 20 mA Input Input LDO 4400 170 1 LDO 2 3 4 5 6 1 Shield Power Out A B...

Page 8: ...damage 1 Switch off the power supply to the device 2 Open the device loosen the 4 screws on the front 3 Plug the module into the slot D SUB connector see figure 4 Tighten the module s fastening screws 5 Connect the sensor cable 6 Close the device by tightening the screws on the front 7 Switch on the power supply NOTICE Moisture ingress The cable glands must be tightly sealed If necessary use suita...

Page 9: ...rnal current input terminals 8 and 10 see terminal plate on page 7 This allows pressure correction of the oxygen measurement Cable e g CA M12 005N485 Color Assignment Terminal number Cable strand LDO 3400 170 module White PWR OUT 6 Brown GND 5 Pink RS 485 B 4 Gray RS 485 A 3 Black Schirm 2 ...

Page 10: ...ue After replacing the sensor or the sensor membrane you must perform an adjustment Procedure Every oxygen sensor has its individual slope Stern Volmer constant cvs and its individual zero point phase angle Both values are altered for example by aging For sufficiently high accuracy of oxygen measurement the analyzer must be regularly adjusted for the sensor data adjustment Calibration Adjustment M...

Page 11: ...certain processes the sensor cannot be removed for calibration Here calibration must be performed directly in the process medium e g by means of a product calibration If there is a temperature difference between the calibration medium and the measured medium you must keep the sensor in the respective medium for several minutes before and after calibration in order to achieve stable measured values...

Page 12: ...ration Measuring Calibration Module A Module B Selecting the measuring module not with 2 optical oxygen sensors K2 contact HOLD is active Module A Calibration Module B Calibration End of calibration or abort Measuring The current output of the selected module is in function check HOLD mode ...

Page 13: ...0 Air Module LDO 3400 170 Module LDO 3400 170 Return 25 6 C Automatic Air Product calibration Sat Zero correction 83 0 Air Automatic Water Select a calibration method Automatic Water Automatic Air Product calibration Sat concentration partial pressure Zero correction When you access calibration the analyzer automatically proposes the previous calibration method If you do not want to calibrate Retu...

Page 14: ...um you must keep the sensor in the respective medium for several minutes before and after calibration Menu Display Selecting a calibration mode Select module The module is in func tion check HOLD mode The assigned current outputs and relay contacts behave as configured BASE Press enter to confirm Select Automatic Water calibration method Remove sensor and immerse it in cal medium air saturated wat...

Page 15: ...ensor in process exit calibration using softkey or enter Automatic Water Cal medium Air sat water 25 6 C 80 3 Air Start Automatic Water Drift check running 25 6 C 80 3 Air Partial pressure Slope correction 214 0 mbar Calibration temperature 025 6 C Calibration pressure 1013mbar Return Display of selected calibration medium Air sat water Enter cal pressure if manual has been configured Start with s...

Page 16: ...es sure constant during calibration If there is a temperature difference between calibration medium and measured medium you must keep the sensor in the respective medium for several minutes before and after calibration Menu Display Selecting a calibration mode Select module The module is in func tion check HOLD mode The assigned current outputs and relay contacts behave as configured BASE Press en...

Page 17: ...ing softkey or enter Calibration medium Air Select First calibration Enter relative humidity e g Ambient air 50 Bottled gas 0 Enter cal pressure if manual has been configured Start with softkey or enter Automatic Air Calibration medium Air 25 6 C 80 3 Air Start Return Relative humidity 0050 Input cal pressure 1013 mbar Automatic Water Drift check running 25 6 C 80 3 Air Partial pressure Slope corr...

Page 18: ...easured at temperature and pressure conditions similar to those of the process Menu Display Product calibration Select module The module is in func tion check HOLD mode The assigned current outputs and relay contacts behave as configured BASE Press enter to confirm Select Product calibration calibration method Sat or Conc p is preset in Parameter setting Cal preset values Press enter to confirm Ca...

Page 19: ...nter Press meas to return to measurement Exception Sample value can be measured on the site and be entered immediately To do so press Input softkey Product calibration Sat Cal medium Product 25 6 C 80 3 Air Cal by taking sample and entering lab value Start Return Product calibration Step 1 Sampling Save the sample value 25 6 C 80 3 Air Input lab value Saturation 80 3 Air Pressure 1014mbar Temperat...

Page 20: ...en you should keep the sensor in the calibration medium eg N2 or sulfite solution until the measured value has stabilized This may well take several minutes After that you can start the calibration process Calibration Return 25 6 C Module PH 3400 035 Module CONDI 3400 051 Info 80 3 Air Module LDO 3400 170 Module LDO 3400 170 Return 25 6 C 80 3 Air Automatic Water Automatic Air Product calibration ...

Page 21: ...vior of the current outputs depends on the parameter setting i e they may be frozen at the last measurement or set to a fixed value The red Alarm LED blinks Measurement operations must not be carried out while the Protos is in the function check HOLD mode as this may put the user at risk due to unexpected system behavior Menu Display Action Open the Parameter Setting menu From the measuring mode P...

Page 22: ...ck icon The functions are released or blocked using the softkey Operator Level Access to all functions which have been released at the Administrator level Blocked functions are displayed in gray and cannot be edited Fig Viewing Level Display of all settings No editing possible Menu selection Return to meas 25 6 C 82 3 Air Select enter Lingua Parameter setting Return 25 6 C 82 3 Air Viewing level A...

Page 23: ...es to Release Open the parameter setting menu Select Operator level passcode 1246 Select Sensor SE 740 for example Now the locked function is displayed in gray and marked with the lock icon Sensor SE 740 Administrator 25 0 C Return Sensor data Cal preset values 82 3 Air Salinity correction Messages Pressure correction Input filter Block Parameter setting Administrator 25 0 C Return Module FRONT 34...

Page 24: ... Air Parameter setting Administrator 25 0 C Return Module FRONT 3400 011 Module BASE 3400 029 82 3 Air System control Sensor SE 740 Administrator 25 0 C Return Pressure correction Sensor data Cal preset values 82 3 Air Salinity correction Messages Input filter Block During parameter setting the analyzer is in function check HOLD mode Current outputs and relay contacts behave as configured BASE mod...

Page 25: ...lect using arrow keys press enter to confirm Menu selection Return to meas 25 0 C Select enter Lingua 83 2 Air HOLD During parameter setting the analyzer is in function check HOLD mode Current outputs and relay contacts behave as configured BASE module Parameter setting Administrator 25 0 C Return Module FRONT 3400 011 Module BASE 3400 029 83 2 Air System control Module PH 3400 035 Sensor SE 740 A...

Page 26: ... details Administrator 20 1 C Zero Sensocheck 100 4 Air Sensor wear Temperature Slope Return Auto Auto Auto Auto Auto Monitoring Auto Sensoface provides information on the sensor condition Great devia tions are signaled Sensoface can be switched off Sensor Monitoring Details The following parameters are moni tored slope zero temperature sensor wear CIP SIP counter O2 measurement with CIP SIP autoc...

Page 27: ... counter Off Individual O2 measurement with CIP SIP Auto Individual Auto off function for O2 measurement When the preset cut off temperature is exceeded the sensor automatically stops the oxygen measurement to extend the life of the membrane The measured oxygen value is frozen the temperature continues to be measured Autoclaving counter Off Individual Cal preset values Product calibration Sat Conc...

Page 28: ...anual xxxx mbar External Pressure during cal Manual xxxx mbar External Salinity correction Entry Salinity Chlorinity Conductivity xx xx g kg 00 00 Messages Messages Saturation Air Off Variable limits Messages Saturation O2 Off Variable limits Messages Concentration Off Variable limits Messages Partial pressure Off Variable limits Note The display may vary depending on the device version ...

Page 29: ...mer constant Zero point phase angle Sensocheck Response time Calibration timer Sensor wear Sensoface Criteria Permitted range 0 0100 0 0350 Permitted range 62 75 Sensor defective first remedy remove and reconnect the sensor or replace the sensor cap 1200 sec Reminds you to calibrate Try replacing the sensor cap Sensocheck The error information is taken from the sensor 10 3 mA 05 02 14 ...

Page 30: ...ing the right softkey displays the message identifier Logbook Administrator Return Erase logbook No Log failure Log warning Yes No Yes Logbook Return 05 13 14 09 50 Measurement active 05 12 14 17 52 Measurement active 05 12 14 17 44 Parameter setting active 05 13 14 09 36 Parameter setting active 05 12 14 17 04 Measurement active 05 12 14 17 40 Wrong passcode 05 12 14 16 53 Diagnostics active Fact...

Page 31: ...ssages Partial pressure Messages Saturation O2 Messages Saturation Air Messages Concentration Messages Administrator 20 1 C 83 4 Air Warning Limit Hi Monitoring Failure Limit Lo Failure Limit Hi Warning Limit Lo 0400 mbar Variable limits 0450 mbar 0100 mbar 0050 mbar Abort OK Message list 22 3 C Fail O2 partial pressure 83 4 Air Fail O2 concentration Abort Diagnostics menu When the Maintenance or ...

Page 32: ...100 50 Assignment of measured values Start 4 mA and end 20 mA Air 20 4 90 80 mA Example 1 Range Air 50 100 Example 2 Range Air 80 90 Advantage Higher resolution in range of interest mA Configuring a Current Output Open parameter setting Enter passcode Select Module BASE Select Output current Module BASE Administrator 19 2 C Return Contact K2 NAMUR HOLD Contact K4 NAMUR Failure Contact K3 NAMUR mai...

Page 33: ...c The process variable is represented by a linear output current curve Note Bilinear characteristic For a bilinear characteristic identical parameters are entered for the two vertices 1st vertex 2nd vertex Trilinear characteristic Two additional vertices must be entered Current Outputs Characteristics Select menu Parameter setting Module BASE Note Function check HOLD mode active ...

Page 34: ... resolution and high values with a low resolution Required Entering a value for 50 output current 100 50 Equation 1 K x 1 Kx E S 2 X50 X50 S K M S E S x S Start value at 4 mA X50 50 value at 12 mA output current range 4 to 20 mA E End value at 20 mA M Measured value Logarithmic output curve over one decade S 10 of maximum value X50 31 6 of maximum value E Maximum value Logarithmic output curve ove...

Page 35: ...is at 63 after the time interval has been reached The time interval can be set from 0 to 120 sec If the time interval is set to 0 sec the current output follows the input Note The filter only acts on the current output and the current value of the second ary display not on the measurement display the limit values or the controller Time interval 0 120 sec Process variable 0 4 20 mA Note For further...

Page 36: ...xample and then checking the values measured Menu selection Return to meas 25 6 C Select enter Lingua 83 3 Air Maintenance Sensor monitor autoclaving counter Note Function check HOLD mode active Setting BASE module Sensor SE 740 25 6 C 83 3 Air Sensor monitor Autoclaving counter Maintenance 25 6 C 83 3 Air Sensor SE 740 Module LDO 3400 170 Module BASE 3400 029 Sensor SE 740 25 6 C 83 3 Air Sensor ...

Page 37: ...ng enter The Diagnostics menu gives an overview of all functions available Functions which have been set as Favorite can be directly accessed from the measuring mode Diagnostic Functions General status information of the measuring system Select menu Diagnostics Menu selection Return to meas 83 1 Air 23 7 C Select enter Lingua Diagnostics Return 23 7 C Set favorite 83 1 Air Logbook Message list Dev...

Page 38: ...ssibilities Module diagnostics Input output status Example Module BASE input output status Device description Return 22 7 C 83 1 Air Digital sensors Module LDO 3400 170 Hardware 1 Software 1 0 Serial number 004711 BASE Module FRONT Module FRONT Return 22 7 C 83 1 Air Display test Keypad test Module diagnostics Module BASE Return 22 7 C 83 1 Air Input output status Set favorite Module diagnostics I...

Page 39: ...r setting System control Function control matrix Sensor Monitor Shows the data provided by the sensor Important function for diagnostics and validation Menu selection Return to meas 23 0 C Select enter Lingua 83 3 Air Sensor SE 740 Return 23 0 C 83 3 Air Cal record Sensor diagram OXY Sensor wear monitor Set favorite Sensor monitor Sensor monitor Return 23 0 C 83 3 Air Ext pressure transmitter Part...

Page 40: ...roduct calibration 71 35 Slope cSv 0 0256 SE 740 1 Sensor network diagram Return 23 0 C 83 3 Air 2 Zero point 1 Slope 4 Cal timer 3 Sensocheck 5 Sensor wear Critical range inner circle Value out of tolerance The tolerance can be modified as required Outer circle Value within tolerance Info Sensor Wear Monitor In addition to the current sensor wear the sensor operating time is displayed The numbers...

Page 41: ... CIP and SIP cycles and correspond ingly increments the counter The user can specify the max number of cycles in the Sensor monitoring details menu and decide whether a message is to be generated when this number is exceeded These data are not stored in the sensor and cannot be reset The number of cycles is shown in the sensor wear monitor of the Diagnostics menu when an individual max value has b...

Page 42: ...tions can be called directly from the measuring mode using a softkey The table on the next page explains how to select favorites Secondary Displays 1 Here additional values are displayed in the measuring mode according to the factory setting When the respective softkey 2 is pressed the process variables measured by the modules plus date or time are displayed In addition you can use the softkeys 2 ...

Page 43: ...an be directly called from the measuring mode Menu Display Selecting a favorite Favorites Menu Diagnostics functions can be called directly from the measuring mode using a softkey The Favorites are selected in the Diagnostics menu Selecting a Favorite Press menu key to select menu Select Diagnostics using arrow keys press enter to confirm Then select the module and confirm by pressing enter 03 09 ...

Page 44: ...ges in plain text Menu selection Return to meas 82 3 Air 25 6 C Select enter Lingua Diagnostics Return 23 7 C Set favorite 82 3 Air Logbook Message list Device description FRONT module Message list Return 25 6 C 82 3 Air Warn BASE module 8 messg Point of meas description Warn Temp alarm HI Saturation Air Alarm HI Fail Wrong ISM sensor Warn Tolerance band Warn Zero range D062 P018 D013 P210 P030 Se...

Page 45: ...16 Temperature Alarm LO_LO FAIL D017 Temperature Alarm LO WARN D018 Temperature Alarm HI WARN D019 Temperature Alarm HI_HI FAIL D020 Concentration range FAIL D021 Concentration Alarm LO_LO FAIL D022 Concentration Alarm LO WARN D023 Concentration Alarm HI WARN D024 Concentration Alarm HI_HI FAIL D025 Part pressure range FAIL D026 Part press Alarm LO_LO FAIL D027 Part press Alarm LO WARN D028 Part p...

Page 46: ...ad Sensoface Membrane wear User defined D080 Range sensor current WARN D081 O2 measurement OFF temp Text D090 Vol range measurement in gases WARN D091 Vol Alarm LO_LO measurement in gases FAIL D092 Vol Alarm LO measurement in gases WARN D093 Vol Alarm HI measurement in gases WARN D094 Vol Alarm HI_HI measurement in gases FAIL D095 ppm range measurement in gases FAIL D096 ppm Alarm LO_LO measuremen...

Page 47: ...Cal Media interchanged Text D205 Cal Sensor unstable Text D254 Module reset Text No Calculation Block OXY OXY Messages Message type H010 AIR Diff Range FAIL H011 AIR Diff Alarm LO_LO FAIL H012 AIR Diff Alarm LO WARN H013 AIR Diff Alarm HI WARN H014 AIR Diff Alarm HI_HI FAIL H015 Temperature Diff Range FAIL H016 Temperature Diff Alarm LO_LO FAIL H017 Temperature Diff Alarm LO WARN H018 Temperature ...

Page 48: ... HI_HI FAIL H090 Vol Diff range measurement in gases WARN H091 Vol Diff Alarm LO_LO measurement in gases FAIL H092 Vol Diff Alarm LO measurement in gases WARN H093 Vol Diff Alarm HI measurement in gases WARN H094 Vol Diff Alarm HI_HI measurement in gases FAIL H095 ppm Diff range measurement in gases FAIL H096 ppm Diff Alarm LO_LO measurement in gases FAIL H097 ppm Diff Alarm LO measurement in gase...

Page 49: ...D022 Out of Specification Concentration Alarm LO D023 Out of Specification Concentration Alarm HI D024 Failure Concentration Alarm LO_LO D025 Failure Partial Pressure Range D026 Failure Partial Pressure Alarm LO_LO D027 Out of Specification Partial Pressure Alarm LO D028 Out of Specification Partial Pressure Alarm HI D029 Failure Partial Pressure Alarm HI_HI D045 Failure Saturation O2 Range D046 F...

Page 50: ...11 User defined SIP Counter D112 User defined Autoclaving Counter D113 User defined Sensor Operating Time D124 Maintenance Required Sensor Date D130 Info SIP Cycle Counted D131 Info CIP Cycle Counted D200 Maintenance Required Temp O2 Conc Sat D201 Maintenance Required Cal Temperature D203 Info Cal Identical Media D204 Info Cal Cal Media Interchanged D205 Info Cal Sensor Unstable D254 Info Module R...

Page 51: ...d Diff Range H021 Failure Concentration Liquid Alarm LO_LO H022 Out of Specification Concentration Liquid Diff Alarm LO H023 Out of Specification Concentration Liquid Diff Alarm HI_HI H024 Failure Concentration Liquid Diff Alarm HI_HI H045 Failure O2 Diff Range H046 Failure O2 Diff Alarm LO_LO H047 Out of Specification O2 Diff Alarm LO H048 Out of Specification O2 Diff Alarm HI H049 Failure O2 Dif...

Page 52: ...or Direct display ofmeasured values from sensor for validation of partial pressure temperature I input Sensoface provides information on the sensor condition Zero slope calibration interval Sensocheck wear Sensor diagram Wear monitor Display of wear parameters Sensor wear sensor operating time autoclaving cycles SIP cycles CIP cycles Current input Pressure range Temperature input Measuring range 1...

Page 53: ...ration Saturation Product calibration Concentration and Product calibration Partial pressure Zero correction Calibration record Recording of Zero slope calibration method with date and time of the last three calibrations and the first calibration User defined at rated operating conditions 1 count plus sensor error Specifications ...

Page 54: ... EMC Emitted interference Interference immunity EN 61326 1 EN 61326 2 3 NAMUR NE 21 Industrial applications EN 55011 Group 1 Class A Industrial applications Lightning protection to EN 61000 4 5 Installation class 2 Rated operating conditions Ambient temperature 20 55 C 4 131 F Relative humidity 10 95 non condensing Transport storage temperature 20 70 C 4 158 F Screw clamp connector Single or stran...

Page 55: ...g OUT PID The corresponding parameters must be set there The minimum current span shall prevent that the resolution limit of the measurement technology 1 count is seen in the current LDO Module air 0 1 O2 0 1 C 10 0 mg l 2 µg l ppm 2 ppb mbar 1 mbar partial pressure vol 0 01 ppm 100 F 10 0 Calculation Block OXY OXY Diff air 0 1 Diff O2 0 1 Diff mg l 10 min 2 0 µg l Diff ppm 10 min 2 0 ppb Diff C 1...

Page 56: ...sfer configuration The complete configuration of a device can be written on a memory card This allows transferring all device settings to other devices with identical equipment exception options and passcodes Parameter set 2 parameter sets A B are available in the device The currently active parameter set is shown in the display Parameter sets contain all settings except sensor type options system...

Page 57: ...lues on the display Display Brightness contrast auto off BASE Module Signal Outputs and Inputs Contacts Output current I1 I2 Separately adjustable current outputs Contact K4 Failure signaling Contacts K3 K2 K1 Separately adjustable relay contacts Inputs OK1 OK2 Optocoupler signal inputs Overview Overview of Parameter Setting 1 Only with Protos II 4400 X Note The menus may vary depending on the dev...

Page 58: ...values Cal saturation Cal concentration Cal timer mg l µg l ppm ppb Salinity correction Entry Salinity Messages Saturation Air Saturation O2 Concentration Partial pressure Pressure correction Ext pressure transmitter Pressure during meas Pressure during cal LDO Module Calibration Menu Automatic Water Automatic Air Product calibration Sat Product calibration Conc Product cal Partial pressure Zero c...

Page 59: ...utoclaving cycles display of maximally permit ted number of cycles when previously entered Diagnostics Menu FRONT Module Module diagnostics Display test Keypad test BASE Module Module diagnostics Input output status LDO Module Module diagnostics Sensor monitor Cal record Sensor diagram Sensor wear monitor Internal function test Shows the values currently measured by the sensor Data of last adjustm...

Page 60: ...n in water 14 B Behavior during messages 35 Bilinear characteristic 33 C Calibration 10 Calibration with sampling 18 Cal record 40 CIP Cleaning in Place 41 Connecting the module 8 Connection assignment 9 Contacts parameter setting 35 Current outputs 32 Current outputs characteristics 33 Current outputs minimum spans 55 D Device description 38 Device firmware 6 Diagnostic messages 44 Diagnostic mes...

Page 61: ...on 6 I Installing the module 8 Intended use 5 L LDO 5 Linear characteristic 33 Lock icon 23 Locking a function 23 Logarithmic output curve 34 Logbook diagnostics 37 Logbook parameter setting 30 M Maintenance 36 Message icons 31 Message list 44 Messages 31 Messages with Protos 3400 45 Messages with Protos II 4400 49 Module compatibility 6 Module installation 8 Index ...

Page 62: ...activating 24 Parameter setting overview 56 Phase angle calibration 10 Product calibration 18 R Rated operating conditions 54 Release softkey function 23 Restore factory settings 30 Returns 2 S Safety information 5 Screw clamp connector 54 Secondary displays 42 Sensocheck 29 Sensoface 26 Sensoface criteria 29 Sensor data 26 Sensor data parameter setting 25 Sensor diagram 40 Sensor monitor diagnost...

Page 63: ...lay 6 SIP Sterilization in Place 41 Specifications 52 Stern Volmer constant calibration 10 Stern Volmer constant permitted range 29 T Table of contents 3 Technical data 52 Terminal plate 7 Trademarks 2 Trilinear characteristic 33 V Variable limits 31 Viewing level 22 W Wiring 9 Z Zero correction 20 Index ...

Page 64: ...s are available for download on our website below the corresponding product description Knick Elektronische Messgeräte GmbH Co KG Headquarters Beuckestraße 22 14163 Berlin Germany Phone 49 30 80191 0 Fax 49 30 80191 200 info knick de www knick de Local Contacts www knick international com 095318 TA 201 170 KNE03 ...

Reviews: