Setting the Sensor Data Parameters
Opening the parameter setting menu
Note:
Function check (HOLD) mode active
Menu Display
Parameter setting
Opening the Parameter Setting Menu
From the measuring mode:
Press
menu
key to select menu.
Select parameter setting using arrow
keys, press
enter
to confirm.
Passcode 1989 (To change passcode:
Parameter setting / System control /
Passcode entry)
Select "Sensor SE 740".
Press
enter
to confirm.
Select using arrow keys,
press
enter
to confirm.
Menu selection
Return to meas
25.0°C
Select:
[enter]
Lingua
83.2 %Air
HOLD
During parameter setting the analyzer
is in function check (HOLD) mode.
Current outputs and relay contacts
behave as configured (BASE module).
Parameter setting (Administrator)
25.0°C
Return
Module FRONT 3400-011
Module BASE 3400-029
83.2 %Air
System control
Module PH 3400-035
Sensor SE 740 (Administrator)
25.0°C
Return
Sensor data
Cal preset values
83.2 %Air
Salinity correction
Messages
Pressure correction
Input filter
Block
Modul LDO 3400-170
Sensor SE 740