15
Menu Display
Automatic calibration
in water
Drift check.
Display during calibration
• Partial pressure
• Calibration temperature
• Calibration pressure
If the signal or the measured tem-
perature fluctuate greatly, the
calibration procedure is aborted.
You must re-start the calibration.
If successful, place sensor in process,
exit calibration using softkey or
enter.
Automatic - Water
Cal medium: Air-sat. water
25.6°C
80.3%Air
Start
Automatic - Water
Drift check running
25.6°C
80.3%Air
Partial pressure
Slope correction
214.0 mbar
Calibration tempe025.6°C
Calibration pressure
1013mbar
Return
Display of selected calibration
medium (Air-sat. water)
Enter cal pressure if “manual” has been
configured.
Start with softkey or
enter
Input cal pressure
1013 mbar
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.