35
Current Outputs: Output Filter
Select menu: Parameter setting/Module BASE/Output current I.../Output filter
Note:
Function check (HOLD) mode active
Time averaging filter
To smoothen the current output, a low-pass filter with adjustable time interval
can be switched on. When there is a jump at the input (100 %), the output level
is at 63 % after the time interval has been reached.
The time interval can be set from 0 to 120 sec. If the time interval is set to 0 sec,
the current output follows the input.
Note:
The filter only acts on the current output and the current value of the second-
ary display, not on the measurement display, the limit values or the controller!
Time interval 0 ... 120 sec
Process variable
0 (4) - 20 mA
Note:
For further BASE module settings (behavior during messages, contacts, opto-
coupler inputs) refer to the user manual of the basic device.