29
Sensoface
Sensoface is a graphic indication of the sensor condition.
The “smileys” provide information on wear and required maintenance of the
sensor (“friendly” - “neutral” - “sad”).
Slope (Stern-Volmer
constant)
Zero point
(phase angle)
Sensocheck
Response time
Calibration timer
Sensor wear
Sensoface Criteria
Permitted range: 0.0100 ... 0.0350
Permitted range: 62° ... 75°
Sensor defective (first remedy: remove and reconnect
the sensor or replace the sensor cap)
> 1200 sec
Reminds you to calibrate
Try replacing the sensor cap
Sensocheck:
The error information is taken from the sensor.
10.3 mA
05/02/14