Parameter Setting Menu
Menu selection
Return to meas
7.00 pH
25.6 °C
Select:
[enter]
Lingua
Parameter Setting
Back
7.00 pH
25.6 °C
Viewing level
Administrator level (All Data) adm
(Operation Data) opl
(All Data) view
Operator level
Parameter Setting
From measuring mode: Press
menu
key to select menu.
Select parameter setting using arrow keys,
press
enter
to confirm.
Administrator level
Access to all functions, also passcode setting. Releasing or
blocking functions for access from the Operator level.
Operator level
Access to all functions which have been released at the
Administrator level. Blocked functions are displayed in gray and
cannot be edited.
Viewing level
Only display, no editing possible!
System Control
Memory card (Option)
Menu only appears when a memory card is inserted and the
corresponding add-on function has been enabled.
Transfer configuration
The complete configuration of a device can be written on a
memory card. This allows transferring all device settings to
other devices with identical equipment (exception: options and
passcodes).
Parameter set
2 parameter sets (A, B) are available in the device.
The currently active parameter set is shown in the display.
Parameter sets contain all settings except:
sensor type, options, system control settings
Up to 5 parameter sets (1, 2, 3, 4, 5) are available when a
memory card (Option) is used.
Function control
Select the functions to be controlled via softkeys and OK inputs
Time/date
Time, date, display format
Meas. point description
Free input of a tag number, can be called from the diagnostics
menu
Release of options
Option activation via TAN
Reset to default
Reset all parameters to factory setting
Passcode entry
Change passcodes
Firmware update
Update the firmware using an Update Card
Logbook
Select the events to be recorded
Overview
Overview of Parameter Setting