19
Menu Display
Product calibration
Product Calibration
Product calibration is performed in
2 steps.
Prepare reference measurement
(e.g. using a portable meter),
start with softkey or
enter
.
Step 2
Lab value has been measured.
When you open the Product calibra-
tion menu again, the display shown
on the left appears:
Enter reference value ("Lab value").
Confirm with “OK”.
Step 1
Take sample.
Save measured value and tempera-
ture at the moment of sampling
(“Save” softkey or
enter
).
Press
meas
to return to measurement.
Exception:
Sample value can be measured on the
site and be entered immediately.
To do so, press “Input” softkey.
Product calibration Sat
Cal medium: Product
25.6°C
80.3%Air
Cal by taking sample
and entering lab value
Start
Return
Product calibration
Step 1: Sampling
“Save” the sample value
25.6°C
80.3%Air
‘Input’ lab value
Saturation
80.3%Air
Pressure
1014mbar
Temperature
+25.6°C
Save
Input
Product calibration
Step 2: Lab value
Input sample lab value
25.6°C
80.3%Air
When changing sensors perform
First cal for statistics!
Lab value
80.0%Air
Abort
OK