TOS6200A/6210
REMOTE and SIGNAL I/O
5-7
5.2.3
Recalling a Panel Memory and Program
Ready status, PM, and STB signals are handled according to the following timing.
Be sure that a READY signal set LOW.
Table 5-2 gives the relation of PM0 to PM7 signals to the panel memory number or
program number to be recalled:
Fig. 5-6
Strobe Signal
Table 5-2
PM Signals and Panel Memory or Program Number to be Recalled
MSD
LSD
MAIN
AUTO
PM7
PM6
PM5
PM4
PM3
PM2
PM1
PM0
H
H
H
H
H
H
H
H
Recalls panel memory 0
Recalls program 0
H
H
H
H
H
H
H
L
Recalls panel memory 1
Recalls program 1
H
H
H
H
H
H
L
H
Recalls panel memory 2
Recalls program 2
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
L
H
H
L
L
H
H
H
Recalls panel memory 98
Recalls program 98
L
H
H
L
L
H
H
L
Recalls panel memory 99
Recalls program 99
READY (OUT)
PM0-PM7 (IN)
STB (IN)
10 ms or more
10 ms or
more
10 ms or
more
5 ms or
more
The next signal input requires an
interval of 10 ms or more.
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...