6-26
GPIB and RS-232C
TOS6200A/6210
6.3.4
Messages Relating to the Tester Status
This subsection describes device messages used to check the status of the tester or
measured values.
FUNCTION (FUN)
Shifts the LCD to each mode (setting screen) or inquires about the current mode
(displayed screen).
Message
Command message:
Fun{0|1|2|3|4}
Query message:
FUN?
Program data
Set value:
0
Test conditions setup screen (MAIN)
1
Program screen (AUTO)
2
Program edit screen (AUTO EDIT)
3
System setup screen (SYSTEM1)
4
Offset screen (OFFSET)
5
Those other than 1 to 4 above (not avail-
able)
Response
Returns the current mode (displayed screen) in response to FUN?.
(Example) When the current mode (displayed screen) is any figure other than 0 to 4
A value of
5
is returned.
IDATA? (IDAT?)
Inquires about the present measured current value.
Message
Query message:
IDAT?
Response
Returns the present measured current value in response to IDAT?
It returns the current value being measured during a test or the measured current
value obtained immediately before when the test was completed.
(Example) When the present measured current value is 25.0 A
A value of
25.0
is returned.
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...