TOS6200A/6210
GPIB and RS-232C
6-19
“OVER LOAD” blinks on the LCD during testing.
If the output power exceeds a maximum power rating, the power limitation safety
function is tripped and protection status occurs.
Change the setting referring to 4.10.3, "Overload Protection (OVER LOAD).”
FREQUENCY (FREQ)
Sets the test frequency or inquires about its set value.
Message
Command message:
FREQ {50|60}
Query message:
FREQ?
Program data
Set value:
50
50 Hz (default)
60
60 Hz
Message
Returns the set value of test frequency in response to FREQ?
(Example) When the test frequency is 50 Hz
A value of
50
is returned.
RJUDGE (RJUD)
Sets ON/OFF of resistance value-based judgment or inquires about its ON/OFF sta-
tus.
Message
Command message:
RJUD {0|1|OFF|ON}
Query message:
RJUD?
Program data
Set value:
0, OFF
Resistance value-based judgment OFF
(Sampled voltage value-based judgment)
1, ON
Resistance value-based judgment ON
(default)
Response
Returns the ON/OFF status of resistance value-based judgment to RJUD?.
(Example) When resistance value-based judgment is ON
“
1
” is returned.
• RJUD ON and VJUD OFF are the same setting.
RJUD OFF and VJUD ON are the same setting.
NOTE
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...