TOS6200A/6210
GPIB and RS-232C
6-21
Response
Returns the lower reference value and ON/OFF status for the lower limit judgement
function in response to LOW?.
(Example) When the lower reference value is 0.5
and the lower limit judgement
is OFF
A value of
0.500,0
is returned.
“UP <= LOW” blinks on the LCD.
If a lower reference value more than the upper reference value is set when lower
limit judgment is ON, the “READY” indication in the upper right of the LCD
changes to blink “UP <= LOW,” informing you that testing cannot begin under the
current conditions. In this case, increase the upper reference value or reduce the
lower reference value.
VLOWER (VLOW)
Sets a sampled voltage value-based lower reference value and ON/OFF of the lower
limit judgment function or inquires about the ON/OFF status of the lower limit
judgment function.
When you perform resistance value-based judgment, use of this setting command
causes sampled voltage value-based judgment to be automatically selected.
Message
Command message:
VLOW <NR2>,{0|1|OFF|ON}
Query message:
VLOW?
•
Program data<NR2>Lower reference value
Set value:
0.01 to 5.40
(default is 0.01)
Resolution:
0.01
Unit:
V
•
Program data{0|1|OFF|ON}Lower limit judgment function
Set value:
0, OFF
Lower limit judgment function OFF (default)
1, ON
Lower limit judgment function ON
(Example) When setting the lower reference value to 3.0 V and the lower limit
judgment function to ON
VLOW 3.0, 1
Response
Returns the lower reference value and the ON/OFF status of the lower limit judg-
ment function to VLOW?.
(Example) When the lower reference value is 3.0 V and the lower limit judgment
function is ON
“
3.00, 0
” is returned.
“UP <= LOW” blinks on the LCD.
If a lower reference value more than the upper reference value is set when lower
limit judgment is ON, the “READY” indication in the upper right of the LCD
changes to blink “UP <= LOW,” informing you that testing cannot begin under the
current conditions. In this case, increase the upper reference value or reduce the
lower reference value.
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...