TOS6200A/6210
Maintenance
7-5
7.6
In Case of Problems
The following problems do not necessarily indicate failures. Please make the fol-
lowing inspections before requesting repairs.
Symptom
Check Item
Chapter/section to be referred to
Page
number
The tester does not start
when the POWER
switch is pressed.
Check that the AC power code is con-
nected.
2.4, “Connecting the Test Leads”
Check that the fuse is blown.
7.3, “Checking and Replacing the Fuse”
The LCD screen does
not appear even when
the POWER switch is
pressed.
Make sure that the contrast adjustment
is correct.
Make sure that ambient temperatures
are not below the recommended opera-
tional limits.
The keys on the front
panel do not work.
Make sure that the key lock function is
off.
Make sure that the tester is not being
controlled remotely via the GPIB or RS-
232C interface.
LOCAL key
The START key does
not function.
Check that no stop signal is being input.
Chapter 5 “REMOTE and SIGNAL I/O”
Check that the tester is not in PROTEC-
TION, PASS, or FAIL status.
Check that the tester is not in the midst
of system setup, interface setup, or pro-
gram editing.
Check that the tester is not currently
saving or recalling information from
panel memory.
Make sure that the double action func-
tion is off.
Check that the "READY" indication is
displayed on the LCD.
Check that “OVER VOLT” is not blink-
ing in the upper right of the LCD.
“OVER VOLT” in 4.9.1, “Setting Output
out of the Operation Range”
Check that “OVER VA” is not blinking
in the upper right of the LCD.
“OVER VA” in 4.9.1, “Setting Output out of
the Operation Range”
Check that “OVER RESI” is not blink-
ing in the upper right of the LCD.
“OVER RESI” in 4.9.1, “Setting Output
out of the Operation Range”
Check that “UP<= LOW” is not blink-
ing in the upper right of the LCD.
4.9.2, “Upper Reference Value <= Lower
Reference Value (UP <= LOW) ”
The fan does not run.
The fan adjusts its rotational rate
depending on the tester's internal tem-
peratures; if the internal temperatures
are sufficiently low, the fan will not run.
If "OVER HEAT" appears on the LCD
and the fan is not running, the tester is
defective.
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...