
2-8
Installation and Preparation for Use
TOS6200A/6210
2.4.3
Other Leads
• If you use a lead wire other than those supplied with the product, select
wires of the nominal sectional areas meeting the test current.
• To avoid generating excessive heat at the connections, use crimp termi-
nals appropriate for the sectional area of the current output line.
Select wires on the basis of the following criteria.
Resistance values with respect to the nominal sectional areas of wires are as follows:
Use lead wires within a total lead length of 10 m or less.
Test Current (I)
Required Nominal Sectional Areas of Wire
I
30 A
5.5 mm
2
or more
30 < I
60 A
14 mm
2
or more
Nominal Sectional Areas of Wire
Resistance Value per Meter
5.5 mm
2
3.5 m
14 mm
2
1.5 m
WARNING
CAUTION
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...