6-20
GPIB and RS-232C
TOS6200A/6210
VJUDGE (VJUD)
Sets ON/OFF of sampled voltage value-based judgment or inquires about its ON/
OFF status.
Message
Command message:
VJUD {0|1|OFF|ON}
Query message:
VJUD?
Program data
Set value:
0, OFF
Sampled voltage value-based judgment
OFF
(Default: resistance value-based judgment)
1, ON
Sampled voltage value-based judgment ON
Response
Returns the ON/OFF status of sampled voltage value-based judgment to VJUD?.
(Example) When sampled voltage value-based judgment is ON
“
1
” is returned.
• RJUD ON and VJUD OFF are the same setting.
RJUD OFF and VJUD ON are the same setting.
LOWER (LOW)
Sets a resistance value-based lower reference value and ON/OFF of the lower limit
judgment function or inquires about the ON/OFF status of the lower limit judgment
function.
When you perform sampled voltage value-based judgment, use of this setting com-
mand causes resistance value-based judgment to be automatically selected.
Message
Command message:
LOW <NR2>,{0|1|OFF|ON}
Query message:
LOW?
•
Program data<NR2>Lower reference value
Set value:
0.001 to 1.200 (TOS6200A) (default is 0.001)
0.001 to 0.600 (TOS6210) (default is 0.001)
Resolution:
0.001
Unit:
OHM
•
Program data{0|1|OFF|ON}Lower limit judgment function
Set value:
0, OFF
Lower limit judgment function OFF
(default)
1, ON
Lower limit judgment function ON
(Example) When the lower reference value is 0.5
and the lower limit judgement
is ON
LOW 0.5, 1
NOTE
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...