TOS6200A/6210
Basic Operation
4-33
4.9.1
Setting Output out of the Operation Range
The testers meet the following test current setting range when the output is within
the maximum rated output and the resistance is such that voltage at the output termi-
nal is 5.4 V or below.
TOS6200A: 3.0 Aac to 31.0 Aac
TOS6210: 6.0 Aac to 62.0 Aac
The following figures show the operation range of the testers.
If a test condition out of this range is set, the LCD displays the cause of it.
a
b
c
OVER RESI
OVER VOLT
OVER VA
(9 A)
(4.5 A)
(60 A, 0.061 Ω)
(40.7 A, 0.132 Ω)
5.4 V
a
b
c
OVER VOLT
OVER VA
(30 A, 0.166 Ω)
(27.7 A, 0.194 Ω)
5.4 V
0
1.2
0.8
1.0
0.2
0.4
0.6
10
25
20
30
Upper reference value, Ω
Test current, A
0
1.2
0.8
1.0
0.2
0.4
0.6
20
50
40
60
Upper reference value, Ω
Test current, A
TOS6200A Operation Range
TOS6210 Operation Range