TOS6200A/6210
Basic Operation
4-27
4.7
Program
A combination of test conditions stored in memory allows you to set 100 programs.
Up to 100 steps may be set as a single program. Note that the number of steps that
can be set for programs is up to 500 in total. For additional information on memory,
see 4.6, "Panel Memory.”
Press the AUTO key to turn on the AUTO key LED and display the program screen
(AUTO).
Program example:
Enter the following settings to create a program like the one above:
00
MEM01-9.9 s
01
MEM02-HOLD
02
MEM03-0s
END
Description of the program example:
The program performs a test of memory 01 in step 00 and then that of memory 02 in
step 01 9.9 seconds after that. Because the interval time of step 01 is set to HOLD,
the test of memory 3 in step 02 does not start unless you press the START switch.
After the START switch is pressed and the test in step 02 is complete, the tester
enters the ready status.
Replacing END at the end of the program with RET instructs the program to repeat
testing from step 00.
AUTO
AUTO
READY
PRG
00 : --UNTITLED--
TOTAL:10
00
MEM01-9.9 s
01
MEM02-HOLD
02
MEM03- 0 s
END
Program name
Step number
Repeat setting
Program number
Running step
Steps waiting to be run
Total number of
steps
Memory 1
Interval of 9.9 s
Memory 2
Interval of HOLD
Memory 3
Interval of 0s
Step 00
Step 01
Step 02
END
Flow of programmed testing
Summary of Contents for TOS6200A
Page 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Page 14: ...1 8 General TOS6200A 6210...
Page 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Page 34: ...3 8 Part Names and Functions TOS6200A 6210...
Page 76: ...4 42 Basic Operation TOS6200A 6210...
Page 134: ...7 6 Maintenance TOS6200A 6210...
Page 150: ...A 8 Appendix TOS6200A 6210...