Triggering and Gating
5
Keysight 53210A User’s Guide
141
A typical measurement using this configuration might return:
+5.00010899135045E-009
With the counter display registering:
5.000 109nsec
(7 digits)
Querying the gate time after sending this command returns:
SENS:FREQ:GATE:TIME?
+1.00000000000000E-003
(1 ms)
Again using
, if a resolution of 7-digits is required the corresponding
gate time can be located and sent directly as:
SENS:FREQ:GATE:TIME 1e-3 // set gate time = 1 ms
External Gate Sources
Gate sources
EXTernal
and
INPut[1]
are external sources.
EXTernal
is the
counter’s rear panel ‘Gate In/Out’ BNC, and
INPut[1]
is the counter’s channel 1
input (front or rear panel - Opt. 201). Note that the
INPut[1]
source is only
available when channel 2 is installed (Option 106 or 115) and is the measurement
channel.