53210A Input Signal Conditioning
106
Keysight 53210A User’s Guide
Input Protection
The maximum input voltage (including any DC offset) allowed with
50
Ω
input
impedance is
±
5.125 Vp. If the input voltage exceeds
~
±
10.0 Vp, the input
protection relay
opens
- changing the input impedance to 1 M
Ω
. (The display and
Impedance softkey will still indicate a 50
Ω
setting however.)
With the relay open, the corresponding channel key flashes until the
input voltage
is removed or reduced below the damage threshold
and the
key is pressed
. This
sets the impedance back to 50
Ω.
To programmatically determine if the protection relay has been
opened
, the
command:
INPut[1]:PROTection?
queries the relay state (0 = relay closed, 1 = relay
open
). With the
input voltage
removed or reduced below the damage threshold
, the protection relay is reset
(closed) with the command:
INPut[1]:PROTection:CLEar
setting the input impedance back to 50
Ω
.
Voltage Overload Bit
The opening of the input protection relay corresponds to
the ‘Voltage Overload’ bit (bit 0) being set in the counter’s Questionable Data
Register group. The bit in the Condition register is cleared by the
INPut:PROTection:CLEar
command or by changing the input impedance to 1
M
Ω
.. The bit in the Event register is cleared by reading the register. See
for more information.
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...