Preparation for Use
44
Keysight 53210A User’s Guide
The minimum frequency at which auto-leveling can occur is set as shown.
Slow sets the minimum frequency for auto-leveling at 50 Hz. Fast sets the
minimum frequency for auto-leveling at 10 kHz.
Refer to
"Threshold Level and Sensitivity"
in
for additional information
on using auto-level.
Keysight 53100 Series Counter Emulation Mode
The 53210A’s emulation mode enables the Keysight 53181A SCPI command set
to be used with the counter.