53210A Error Messages
A
Keysight 53210A User’s Guide
237
+820
Model and Serial Numbers not restored
Following replacement of the P500 processor (controller) board or
main measurement board, the model number and/or serial
number do not match or have not been re-stored. Re-store the
model number and serial number as prompted from the front
panel. Refer to the
53210A Service Guide
for additional
information.
+821
Controller and measurement board
model numbers do not match
+822
Controller and measurement board serial
numbers do not match
Calibration Errors
+701
Calibration error;
security defeated
Shorting the calibration secure jumper (CAL ENABLE) when
turning ON the instrument will cause this error to occur indicating
the security password has been overwritten. See “Resetting the
Security Code to a Null” in the
53210A Service Guide
for details.
+702
Calibration error;
calibration memory is secured
A calibration cannot be performed when calibration memory is
secured. See “To Unlock The Counter For Calibration” and “To
Lock the Counter” in Chapter 2 of the
53210A Service Guide
for
more information. Use the CAL:SEC:STAT ON command to enter
the security code using the remote interface.
+703
Calibration error;
secure code provided was invalid
The security code specified was invalid.
+704
Calibration error;
secure code too long
The security code specified is greater than 12 characters.
+705
Calibration error;
calibration aborted
An in-progress calibration was aborted (stopped) by the user.
+706
Calibration error;
provided value out of range
A calibration value was entered that was outside the expected
range of valid inputs.
+707
Calibration error;
computed correction factor out of range
If this error occurs, cycle power and run the instrument self-test.
Verify the calibration setup (instrument settings, connections) and
recalibrate the counter. Contact Keysight if there is a self-test
failure or the unit fails to calibrate.
+711
Calibration error;
calibration string too long
The message to be stored in calibration memory is greater than 40
characters.
+712
Calibration failed
A hardware failure occurred. Contact Keysight.
+713
Channel 2 calibration signal not
detected
The calibration source signal is not connected to channel 2.
Table A-1
53210A Error Message Descriptions.
Code
Message
Description
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...