A
53210A Error Messages
232
Keysight 53210A User’s Guide
-178
Expression data not allowed
An expression was encountered at an instance where it is not
allowed by the instrument.
Execution Errors
-203
Command Protected
A password-protected command or query could not be executed
because the command was disabled.
-213
INIT ignored
INITiate:IMMediate was received while the counter is currently
initiated.
-221
Settings conflict
The statement appended to the ‘Settings conflict’ message
describes the conflict and how it was resolved.
-222
Data out of range
The parameter value received was outside the range for the given
command.
-223
Too much data
The block, element, or string parameter received is valid, but
memory limitations prevent command execution.
-224
Illegal Parameter Value
An exact value from a list of choices was expected.
-225
Out of memory;
measurement data overrun
Internal buffers are full as data is taken at a faster rate than can be
processed. Decreasing the reading rate (gate time) can prevent
this error.
-230
Data corrupt or stale
Attempting to retrieve data following a reset or change to the
counter configuration.
-240
Hard ware error
Most hardware errors are detected at instrument power-on. If a
hardware error occurs, cycle power again. If the error remains,
contact Keysight.
-241
Hard ware missing
The function specified requires optional channel 2. The channel is
either missing or incorrectly installed.
-250
Mass storage error:
file read/write error
An error occurred reading from, or writing to a file in internal flash
memory or on a USB memory device. File space may have been
exceeded or the USB device removed before the operation was
complete.
-252
Missing Media
Internal flash memory or the USB device was not found during a
file operation.
Table A-1
53210A Error Message Descriptions.
Code
Message
Description
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...