4
53210A Input Signal Conditioning
108
Keysight 53210A User’s Guide
Setting the Probe Factor
For better access to test points on the device under test (DUT), Keysight
recommends its
N2870A
,
N2873A
, and
N2874A
Passive Probes for use with the
53220A/ 53230A counter. The 35 MHz (BW) N2870A 1:1 probe and the 500 MHz
N2873A and 1.5 GHz N2874A 10:1 probes are popular accessories commonly
used with Keysight oscilloscopes.
When using probes with the counter, specifying a probe factor references and
displays
threshold levels
and
p-p levels
relative to actual levels on the DUT,
rather than to the level at the channel input. The probe factor is set with the
command:
INPut[1]:PROBe {<factor>|MINimum|MAXimum|DEFault}
INPut[1]:PROBe? [{MINimum|MAXimum|DEFault}]
(query form)
Probe
factor
settings are
1
for a 1:1 probe (N2870A), and
10
for a 10:1 probe
(N2873A or N2874A). When set to 1, the counter ranges are 5V and 50V. When
set to 10, the counter ranges are 50V and 500V.
Probe Input Compatibility
The input compatibility of the N2870A and N2873A
probes is 1 M
Ω
respectively. The compatibility of the N2874A is 50
Ω
. Each is
compatible with the counter’s 20 pF input capacitance. When using a probe, the
counter input impedance should be set accordingly (1M
Ω
, 50
Ω
).
CONFigure
and
MEASure
do not change the probe factor setting. Following a reset
(
*RST
) or an instrument preset (
SYSTem:PRESet
or
Preset
key), the probe factor is
set to 1.
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...