4
53210A Input Signal Conditioning
120
Keysight 53210A User’s Guide
The slope (edge) of the input signal on which the threshold level occurs is
specified with the command:
INPut[1]:SLOPe {POSitive|NEGative}
INPut[1]:SLOPe?
(query form)
POSitive
- the trigger point occurs on the positive (rising) edge. The reset point
occurs on the negative (falling) edge (
NEGative
- the trigger point occurs on the negative edge, with the reset point
occurring on the positive edge.
CONFigure
and
MEASure
do not change the slope setting. Following a reset (
*RST
)
or instrument preset (
Preset
), the positive (rising) slope is selected.
Input Slope Example
//set the threshold slope
CONF:FREQ 1E6, 0.1, (@1)
// measure 1 MHz signal
SYST:ALEV:FREQ 10E3
// set minimum auto-level frequency
INP:IMP 50
// set impedance to 50 ohms
INP:RANG 5
// set range to 5V
INP:COUP AC
// set AC coupling
INP:LEV:REL 70
// 70% threshold level
INP:SLOP POS
// trigger on positive slope