53210A Math, Graphing, and Data Logging
6
Keysight 53210A User’s Guide
153
Following a reset (
*RST
) or instrument preset (
SYSTem:PRESet
), a reference value
of 0.0 is set and automatic reference is enabled.
From the front panel, pressing
Get New Reference
takes a reference
measurement on the first trigger received after the key is pressed. A reference can
be entered manually by selecting
Ref Value
and using the knob or
SHIFT
ed
numeric keys.
Reference Example
//determine the percentage difference of 100 frequency
//measurements to a reference value of 50000.000
CONF:FREQ 50E3,(@1)
SAMP:COUN 100
// take 100 readings
CALC:STAT ON
// enable CALCulate1 subsystem
CALC:SCAL:STAT ON
// enable scaling
CALC:SCAL:FUNC PCT
// select scale function (PCT)
CALC:SCAL:REF 50.000E3
// set reference value
INIT
Scale Gain and Offset
The function
CALCulate[1]:SCALe:FUNCtion SCALe
performs an
Mx-B
operation
on each reading where
M
is the gain value,
x
is the reading, and
B
is the offset.
The gain value used in the equation is set with the command:
CALCulate[1]:SCALe:GAIN {<gain>|MINimum|MAXimum|DEFault}
CALCulate[1]:SCALe:GAIN? [{MINimum|MAXimum|DEFault}]
(query form)
gain
values can range from:
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to +1.0E+15
The default
gain
is
1.0
and is the value set following a reset (
*RST
) or instrument
preset (
SYSTem:PRESet
).
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...