53210A Math, Graphing, and Data Logging
184
Keysight 53210A User’s Guide
There are 1,000 readings in each decimation group in
readings/100 points displayed). The number of readings in each group changes if
Show All
is pressed and readings continue to be taken.
When a marker is moved to a decimation point, the number of the
first
reading in
the corresponding group is displayed. This number should be noted if it is
necessary to view individual readings within a group.
Viewing Readings Within a Decimation Group
Decimated reading values that
are grouped and represented by a single decimation point can be viewed
individually using markers and the trend chart’s
Pan
control.
1
Select a marker and using the knob, move the marker to the decimation point
with the reading level of interest. The reading number corresponding to the
marker position is the
first
reading in the group. This may not be the maximum
or minimum value within the group.
2
Select ‘
Pan
’ and using the
[Shift]
ed numeric keys, enter the number of the first
reading. The reading indices show a span of 100
readings
, beginning with the
first reading in the group.
3
Select
Marker1
or
Marker2
and using the arrow keys, highlight the rightmost
(1's) place in the position window. Rotate the knob to move the marker one
reading at a time. The reading value and number are shown as the marker is
advanced.
For decimation groups with more than 100 readings, select ‘
Pan
’ and enter the
next reading number after the upper reading index (
) to view the next
100 readings.
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...