![Keysight 53210A User Manual Download Page 139](http://html1.mh-extra.com/html/keysight/53210a/53210a_user-manual_1961097139.webp)
Triggering and Gating
5
Keysight 53210A User’s Guide
139
Setting the Gate Time
A measurement gate time can be specified directly with the command:
[SENSe:]FREQuency:GATE:TIME {<time>|MINimum|MAXimum|DEFault}
[SENSe:]FREQuency:GATE:TIME? {MINimum|MAXimum|DEFault}
(query form)
CONFigure
and
MEASure
automatically set the gate source to
TIME,
and the gate
time according to their optional
expected value
and
resolution
parameters.
Specifying these parameters or using default values with these commands
simplifies
counter programming. However, setting/changing the gate time
separate from the
CONFigure
/
MEASure
command allows you to change
only
the
gate time while preserving other configuration settings.
Table 5-2
Resolution and Gate Time.
Gate Time (53210A)
(T
ss
= 100 ps)
Resolution (Digits)
Reciprocal
Calculating Digits and Gate Time
1 msec
7
Digits as a function of gate time:
Digits = Log
10
(gate time/T
ss
)
Digits as a function of expected value
(CONFigure, MEASure)
Digits = Log
10
(
expected value)
-
Log
10
(
resolution
)
Gate time as a function of digits:
gate time = (10^digits) * T
ss
10 msec
8
100 msec
9
1 sec
10
10 sec
11
100 sec
12
1000 sec
13
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...