U3022AE10
Control Lines
External Voltage Supply Configuration
Figure
48
illustrates an example of the connection with an external DC power
supply. Connect the positive and negative voltage supply from the external power
supply to the positive input (pin 13) and the negative input (pin 14). Connect the
power supply DC ground to pin 15.
Turning
on
the Test Set using an External Power Supply.
1.
Turn
on
the test set.
2.
Connect the DUT.
3.
Turn
on
the external power supply.
Turning
off
the Test Set using an External Power Supply.
1.
Turn
off
the Power Supply.
2.
Turn off the test set.
3.
Disconnect the DUT.
Figure
48
Test Set to the DUT and External DC Power Supply
8
1
15
14
13
TO DUT (Only required Lines)
GND
+
Gnd
Ͳ
Gnd
DC Power Supply
12
13
14
15
1
2
8
2 to 5 Vdc
TEST SET
Line 1
Line 2
Line 8
GND
Positive: 0 to 5 Vdc
Negative:
Ͳ
5 to 0 Vdc
NC
9
(Rear Panel View)
Keysight U3022AE10 User's and Service Guide