U3022AE10
General Specifications
Frequency Range and
Power Levels
Table 5
Recommended Maximum Power Levels
U3022AE10 Test Port RF Power Levels:
PORT 3-10
+27 dBm
U3022AE10 Access Ports:
SOURCE OUT
+27 dBm
CPLR ARM
+27 dBm
CPLR THRU
+27 dBm
RCVR A IN and B IN
+12 dBm
Damage Power Levels to U3022AE10 Access and Test Ports:
Max Level to Port 1 & 2 Test Ports
+30 dBm
Refer to your analyzer’s specifications to determine the maximum input power levels
for the access and test ports, or to optimize the power levels in the receivers.
Damage and maximum levels are not necessarily the optimum level.
When the test set is configured with your network analyzer, the system frequency range
of operation will be determined by the frequency range limits of this combination.
It is recommended that you do not operate components near damage levels (+30
dBm). The power levels must be 3 dB below maximum level to ensure no damage.
Do not apply any DC voltage to the Access or Test Set ports.