U3022AE10
Troubleshooting the Test Set
Equipment Setup
1.
Turn
on
the test set and the analyzer.
2.
Preset
the analyzer and set it to
Standalone Mode
. Select
Utility
>
System
>
Configure
>
Multiport Capability
. In the dialog box select
Restart as a standalone
PNA
>
OK
.
3.
Connect the RF cables to the analyzer’s Ports 1 and 2. Connect the cables
together using a 3.5 mm adapter.
4.
Configure the analyzer to measure S21 and normalize the response trace.
5.
Set the analyzer to
Interface Control Mode
: Select
Channel > Hard
ware Setup >
More > Interface Control…
and click
Enable Interface Control box
.
The <addrs>.<data> entries noted in the following Test Instructions table will be used
to configure the RF switches for this testing. After making your entry select <OK> to
execute the command, to return back for further entries, select Interface Control on
the PNA display.
Table
19
RF Signal Path Insertion Loss (S21)
RF Path Description
Signal
Insertion Loss (typical)
Source Out to Ports 3-12
Source
A/B IN to Ports 3-12
Receiver
Source Out to CPLR THRU
Source
A/B IN to CPLR ARM
Receiver
Keysight U3022AE10 User's and Service Guide
81