U3022AE10
Troubleshooting the Test Set
Connect the RF Flex cables to the Source Out and CPLR THRU port indicated
in Table 2
2
and Table 2
3
, the expected results should be similar to Figure
68
.
Figure
68
Source Out to CPLR THRU or A/B IN to CPLR ARM, Standard
Table 2
2
Source Bypass Signal Path Test Instructions
Path #
Test Port
Group
RF Path
Description
Control Mode
<Address>.<Data>
Path
Components
Insertion Loss
(typical)
1
Left Side
Source Out
to
CPLR THRU
0.0
S10
2
Right side
Source Out
to
CPLR THRU
16.0
S20
Table 2
3
Receiver Bypass Signal Path Test Instructions
Path #
Test Port
Group
RF Path
Description
Control Mode
<Address>.<Data>
Path
Components
Insertion Loss (typical)
1
Left Side
A IN to
CPLR ARM
0.0
S11
2
Right side
B IN to
CPLR ARM
16.0
S21
Figure
68
Figure
68