U3022AE10
Cal Kit Operational Check
Cal Kit Operational Check Procedure
The sequence of this procedure is very important and must be followed or the
performance accuracy and results may vary from the reference plots provided.
In this procedure (when using with the 4-Port PNA-L or PNA-X System) Ports 13 and
14 are not connected to the test set, therefore not verified.
If you suspect that your 12 or 14-Port configuration is not operating properly, ensure
that all front RF jumper interconnect cables are correctly attached.
Preparing the Network Analyzer
1.
Connect the test set to the analyzer using the interconnect cables as shown in
“Hardware Lock-link Kit Installation (U3021-60001)” on page
Lock-link Kit Installation (U3021-60002)” on page
2.
Turn
on
the test set.
3.
Select
Response > Cal > Manage Cals > Cal Set
-OR-
(Cal Sets & Cal Kits > Cal Set...)
.
Delete or rename any Cal Sets
titled “999.1” thru
“999.12” (12-Port), although it is unlikely that you will find Cal
Sets with these
names.
4.
Verify that the network analyzer is in Multiport mode. See the bottom of the
measurement window, similar to
.
a.
If only four S-Parameters are listed, select
Utility
> System > Configure >
Multiport Capability
-OR-
(Instrument > Setup > External Hardware > Multiport > Multiport
Configuration...)
. Select
Restart as multiport PNA with this testset
and
select
U3022AE10
(12-Port) from the drop-down menu >
OK
5.
Press
[Preset]
.
6.
Verify that the
[Start Frequency]
is set to
[300 kHz or 10 MHz]
.
7.
Set the
[Stop Frequency]
to
[20 GHz or 26.5 GHz]
.
8.
Select
[Power]
>
Power Level
and enter
[0 dBm]
.
9.
Select
Response
>
Avg
>
IF Bandwidth
>
100 Hz > OK
.
10.
Select
Stimulus
>
Sweep
>
Number of Points
>
401
.
11.
Connect the ECal module to an available USB port on the front or rear panel. This
procedure assumes you are using a ECal. If you are not, see
Verification Procedure” on page 6
12.
Allow the ECal module, test set and analyzer to warm up for a minimum of 30
minutes.