U3022AE10
Troubleshooting the Test Set
Figure 6
5
Source Out to Ports 3-12 Path Response
Table 2
0
Source Signal Path Insertion Loss Test
Connect the RF Flex cables to the Test Port and Source Out port indicated
in Table 2
0
, the expected results should be similar to Figure 6
5
.
Source Signal Path Test Instructions
Path #
RF Path
Description
Control Mode
<Address>.<Data>
Path
Components
Insertion Loss
(typical)
1
Source Out to Port 3
a
0.1
P3 CPLR, S10
Figure 6
5
2
Source Out to Port 5
0.2
P5 CPLR, S10
3
Source Out to Port 7
0.3
P7 CPLR, S10
4
Source Out to Port 9
0.4
P9 CPLR, S10
5
Source Out to Port 11
0.5
P11 CPLR, S10
6
Source Out to Port 4
16.1
P4 CPLR, S20
7
Source Out to Port 6
16.2
P6 CPLR, S20
8
Source Out to Port 8
16.3
P8 CPLR, S20
9
Source Out to Port 10
16.4
P10 CPLR, S20
10
Source Out to Port 12
16.5
P12 CPLR, S20
a. Use the Source Out port (on the left) associated with the odd test ports.